Simmons, Owen, De Wilde, Philippe, Podoleanu, Adrian (2026) Image-driven OCT calibration from a single B-scan. Biomedical Optics Express, 17 (9). ISSN 2156-7085. (doi:10.1364/BOE.608117) (KAR id:115912)

PDF Publisher pdf
Language: English


Download this file
(PDF/6MB)
[thumbnail of boe-17-9-4778.pdf]
Preview
Request a format suitable for use with assistive technology e.g. a screenreader
PDF Author's Accepted Manuscript
Language: English

Restricted to Repository staff only

Contact us about this publication
[thumbnail of Image_Driven_OCT_Calibration_from_a_Single_B_Scan (2).pdf]
Official URL:
https://doi.org/10.1364/BOE.608117

Abstract

Optical coherence tomography (OCT) requires accurate calibration of wavenumber sampling and dispersion compensation to achieve optimal axial resolution. Conventional calibration methods rely on dedicated mirror measurements acquired at multiple optical path differences, requiring dedicated acquisitions, careful alignment, and additional acquisition time. We present a physics-informed optimisation framework that recovers OCT calibration directly from a single routine B-scan. Within the complex leader-follower reconstruction framework, the method jointly estimates wavenumber remapping and dispersion compensation functions by optimising image quality in reconstructed OCT data. Experiments on a swept-source OCT system demonstrate that the recovered calibration functions produce image quality comparable to conventional mirror-based calibration while requiring only a single acquired image and 30–60 seconds of optimisation. Samples containing depth-distributed structure recover calibration functions closely matching the underlying system calibration, whereas structurally limited samples produce image-specific effective calibration within the observed depth range. These results demonstrate the feasibility of rapid software-defined OCT calibration directly from imaging data, reducing dependence on dedicated calibration measurements, and simplifying OCT system operation.

Item Type: Article
DOI/Identification number: 10.1364/BOE.608117
Additional information: For the purpose of open access, the author(s) has applied a Creative Commons Attribution (CC BY) licence to any Author Accepted Manuscript version arising.
Subjects: Q Science > QC Physics > QC355 Optics
Institutional Unit: Schools > School of Engineering, Mathematics and Physics > Physics and Astronomy
Schools > School of Natural Sciences > Biosciences
Former Institutional Unit:
There are no former institutional units.
Funders: Medical Research Council (https://ror.org/03x94j517)
National Institute for Health Research (https://ror.org/0187kwz08)
Engineering and Physical Sciences Research Council (https://ror.org/0439y7842)
Depositing User: Owen Simmons
Date Deposited: 18 Aug 2026 08:08 UTC
Last Modified: 11 Sep 2026 15:45 UTC
Resource URI: https://kar.kent.ac.uk/id/eprint/115912 (The current URI for this page, for reference purposes)